Skip to main content

Why Inspection Can Fail After an Instrument Passes Calibration

Calibration Troubleshooting

Investigate failed inspection results by checking instrument suitability, test points, uncertainty, setup, environment, method, and product condition.

Your Calibration Certificate Passed. Why Is the Part Still Failing Inspection?

The instrument has a valid calibration certificate. The calibration result is acceptable. Yet, the part still fails inspection.

This can seem contradictory when calibration and inspection are viewed as the same. They are not.

Calibration checks how measurement standards match what a measuring system shows under specific conditions. Inspection checks if an item meets set requirements.

When results conflict, review the entire measurement process before blaming the instrument.

Calibration and Inspection Answer Different Questions

Calibration and product inspection serve different purposes.

Calibration connects the values from measurement standards to what the measuring system shows in specific conditions.

Inspection uses measurements or other evidence to see if an item meets requirements.

A calibration certificate shows an acceptable result. But, it doesn’t ensure that every measurement taken during production or inspection will be acceptable.

 JCGM VIM: 

Calibration

Check the instrument

Start by confirming that the instrument used for the inspection is the instrument identified in the calibration records.

  • Confirm the instrument identification.
  • Check the calibration status.
  • Review the relevant calibration results.
  • Check the measurement range.
  • Check the instrument condition.
  • Review intermediate checks where applicable.

Also consider whether the instrument was configured or used differently from the conditions covered by the applicable procedure.

A calibration record should be interpreted in relation to the measurement task for which the instrument is being used.

Review the inspection method

The next step is to examine how the part was measured.

Review the details that apply to the particular measurement:

  • Measurement location
  • Positioning
  • Alignment
  • Reference surfaces
  • Contact conditions
  • Fixtures
  • Measurement sequence
  • Operator procedure

A valid calibration record does not remove variation introduced by the measurement application.

If two people use the same instrument differently, or if the part is positioned differently between measurements, the results may not be directly comparable.

Check the specification

Confirm the requirement against which the part is being inspected.

Record the actual tolerance or acceptance requirement. Check that the correct revision of the specification, drawing, customer requirement or documented procedure is being used.

Also confirm that the measured characteristic is the one specified by the requirement.

Do not assume that an expected value is the same as the actual acceptance criterion.

Consider measurement uncertainty

Measurement uncertainty can become important when an inspection result is close to a specification boundary.

Measurement uncertainty describes the dispersion of quantity values attributed to a measurand based on the available information.

If the result is close to the limit, identify the applicable decision rule before deciding whether the item conforms.

Do not apply a universal guard band or acceptance formula. Use the decision rule specified by the applicable standard, specification, customer requirement or documented procedure.

JCGM VIM: Measurement uncertainty

Investigate the part and process

The measurement may be showing a genuine change in the component.

If the inspection result differs from previous results, compare the available measurement records with relevant production information.

Depending on the manufacturing process, useful records may include equipment settings, tooling changes, maintenance activity, material information or other documented process changes.

Do not assume that a production change caused the measurement result. Treat the connection as a finding only when the available evidence supports it.

Likewise, do not blame the measuring instrument simply because the part failed inspection.

Keep the evidence together

A measurement investigation is easier to review when the relevant records are kept together.

Document:

  • Instrument identification
  • Calibration status
  • Calibration results where relevant
  • Measured value
  • Measurement method
  • Measurement conditions
  • Applicable specification
  • Measurement uncertainty where relevant
  • Decision rule
  • Final decision

This record allows another qualified person to understand what was measured, how it was measured and why the resulting decision was made.

Inspection investigation checklist

  1. Confirm the instrument identification.
  2. Confirm calibration status.
  3. Review the relevant calibration result.
  4. Check the measurement range.
  5. Confirm the inspection method.
  6. Check the measurement location.
  7. Review positioning and alignment.
  8. Check reference surfaces, fixtures and contact conditions where applicable.
  9. Confirm the applicable specification and tolerance.
  10. Consider measurement uncertainty for borderline results.
  11. Identify the applicable decision rule.
  12. Review relevant part and process records.
  13. Document the evidence and final decision.

Continue the investigation

For the broader question of why a calibrated instrument can still produce an unexpected measurement, read:

Your Instrument Is Calibrated. Why Can the Measurement Still Be Wrong?

If two calibrated instruments give different readings, read:

Two Calibrated Instruments Give Different Readings. Which One Should You Trust?

If recalibration did not remove the problem, read:

The Instrument Was Recalibrated, but the Measurement Problem Remains

If the result is just outside the tolerance, read:

A Measurement Is Just Outside Tolerance. Should You Reject the Part?

Comments

Popular posts from this blog

Vernier Caliper Calibration Procedure, Error and Uncertainty

Dimensional Calibration Review caliper calibration for external, internal, depth, and step measurements, including reference standards, error, repeatability, and uncertainty. Contents What Is Vernier Caliper Calibration? Why Does Vernier Caliper Calibration Matter? Vernier Caliper Types Covered by Calibration Current Indian Standard for Vernier Calipers What Is Checked During Vernier Caliper Calibration? Vernier Caliper Calibration Procedure: Step by Step External, Internal and Depth Measurements Measurement Uncertainty in Vernier Caliper Calibration What Should a Vernier Caliper Calibration Certificate Contain? What Does Calibration Traceability Mean? NABL Vernier Caliper Calibration: What Should You Verify? How to Read a Vernier Caliper Calibration Certificate Dimensional Metrology •  India This guide explains what is checked during Vernier caliper calibration, what a calibration certificate should contain, and how to assess a laboratory's technical scope. Key p...

Dimensional Calibration for Manufacturing and Quality Teams

Dimensional Calibration Build a dimensional calibration plan around intended use, suitable references, environmental control, test points, uncertainty, records, and result review. Contents Dimensional Calibration: A Practical Guide to Dimensional Measurement What Is Dimensional Calibration? Why Is Dimensional Measurement Important? What Is Dimensional Measurement? Common Dimensional Measuring Instruments How Does Dimensional Measurement Work? Dimensional Measurement Procedure Why Measurement Direction and Alignment Matter Why Temperature Matters in Dimensional Measurement Factors That Can Affect Dimensional Measurement How Is Dimensional Calibration Performed? Calibration and Verification of Dimensional Instruments Dimensional Metrology & Calibration Guide Dimensional Calibration: A Practical Guide to Dimensional Measurement Understand how dimensional measurements are made, what can influence them, how calibration is performed, and what to check before selecting a...

Pipette Calibration Checks Before Trusting Dispensed Volume

Mass Balance and Volume Review pipette identity, selected volume, tips, test method, environment, balance, results, uncertainty and decision rule before trusting a dispensed volume. Contents Direct answer Calibration, routine testing and adjustment Identify the complete pipetting system Confirm the measurement method Control the balance, liquid and environment Read systematic and random effects separately Apply the correct acceptance requirement Review the calibration certificate Build a useful routine check Direct answer A calibration check label does not set up that every quantity brought with the aid of a pipette is suitable for every method. Consider the result only after confirming the pipette configuration, decided on test volumes, recommendations, measurement approach, environmental controls, dimension uncertainty and attractiveness requirement. ISO 8655-2:2022 covers me...