NABL & ISO/IEC 17025:2017 Calibration Reference Portal
Official Lab Services: multitek.in
Solutions for: Quality Heads & Auditors Shop Floor Inspectors Tool Room Technicians Engineering Students

NABL 138 Air Quality Monitoring Equipment Calibration 2026: Gas Analyzers, PM Monitors and Traceability

NABL 138 Air-Quality Monitoring Equipment Calibration: Requirements for Gas Analyzers and PM Monitors

Updated: 16 September 2026

NABL 138 sets specific criteria for laboratories that calibrate air-quality monitoring equipment. It applies to gas analyzers used for pollutants such as NOx, CO, SO₂, O₃, ammonia, and benzene, as well as PM2.5 and PM10 analyzers.

The NABL Accreditation Documents register lists NABL 138 as Issue 01, originally issued on 22 January 2020, with Amendment 02 dated 3 November 2025. This remains the current published version as of September 2026.

NABL 138 works alongside ISO/IEC 17025, the international standard for testing and calibration laboratory competence. ISO confirms that ISO/IEC 17025:2017 was reviewed and confirmed in 2023 and remains current.

What is NABL 138?

NABL 138 is titled Specific Criteria for Air Quality Monitoring Equipment Calibration Laboratories.

The document adds equipment-specific requirements to the general laboratory competence framework established by ISO/IEC 17025. It addresses the reference standards, equipment, calibration methods, environmental controls, personnel competence, traceability, and measurement uncertainty needed for air-quality analyzer calibration.

Why is a separate document necessary?

Because an air-quality monitoring station does not measure pollution merely because a number appears on its screen. Every result depends on a chain of measurements.

A gas analyzer relies on known reference concentrations, controlled gas flow, dilution equipment, instrument response, and stable operating conditions. A particulate monitor depends on sampling flow, its particle-measurement system, temperature and pressure corrections where applicable, and the condition of the instrument.

If one part of that chain is wrong, the reported concentration may also be wrong.

Why air-quality equipment calibration matters

Air-quality measurements can shape public-health guidance, pollution-control decisions, industrial compliance, and environmental policy. Poor calibration can make a functioning instrument produce results that appear credible but are biased.

That creates problems far beyond a single monitoring station. If two stations respond differently to the same pollutant concentration, comparisons between locations may become misleading. Trends can appear where none exist, and genuine changes can be missed.

Calibration connects an analyzer’s response to suitable reference standards. It helps laboratories and monitoring organizations determine whether the equipment is measuring correctly, identify bias, and estimate uncertainty in the result.

Which instruments does NABL 138 cover?

NABL 138 divides air-quality monitoring equipment into two broad groups.

Equipment groupExamples
Gas analyzersNOx/NO₂, CO, SO₂, O₃, ammonia, and benzene analyzers
Particulate-matter analyzersPM2.5 and PM10 analyzers

A laboratory may seek accreditation only for activities it can perform competently. The analyzer type, calibration range, facility mode, measurement capability, and relevant method must be reflected in the laboratory’s approved scope.

Customers should never assume that general NABL accreditation covers every air-quality instrument. They should check the laboratory’s current scope before ordering calibration.

Permanent, on-site, and mobile calibration

Air-quality analyzer calibration may be carried out through permanent, on-site, or mobile facilities, depending on the laboratory’s approved scope and technical capability.

This flexibility matters because many analyzers operate inside continuous monitoring stations. Removing them may interrupt data collection and expose sensitive equipment to transport risks.

On-site calibration solves some practical problems, but it introduces others. The laboratory must still control the setup well enough to produce valid results. Relevant conditions may include:

  • ambient temperature and humidity;
  • station ventilation;
  • electrical supply;
  • gas-line configuration;
  • reference-equipment transport;
  • available setup space;
  • stabilization time;
  • field records and data capture.

The laboratory should determine which conditions can affect the calibration and document them accordingly.

How are gas analyzers calibrated?

A gas analyzer is calibrated by exposing it to known concentrations of the target gas and comparing its indicated values with the reference values.

For example, an SO₂ analyzer may be challenged with certified sulfur-dioxide gas at several controlled concentrations. The laboratory records the analyzer’s response and evaluates its performance across the required range.

A proper calibration may include:

  • zero gas;
  • certified reference gas;
  • a gas-dilution or calibration system;
  • controlled gas flow;
  • adequate stabilization time;
  • multiple calibration points;
  • repeatability measurements;
  • zero and span behavior;
  • measurement uncertainty.

A zero check and a single span check can help track routine performance, but they do not necessarily provide the same evidence as a complete traceable calibration.

Reference gases and traceability

Reference gases form a critical part of gas-analyzer calibration. Depending on the method, laboratories may use certified span gases or gas mixtures from competent reference-material producers or national metrology institutes.

Yet the cylinder is only one part of the process.

When a laboratory uses a diluter to generate lower concentrations, the concentration delivered to the analyzer depends on the entire system. Relevant factors include:

  • the certified reference-gas value and uncertainty;
  • mass-flow controller or flow-meter performance;
  • the gas-flow ratio;
  • zero-gas purity;
  • leaks;
  • system stability;
  • adsorption or reaction within tubing and fittings;
  • environmental conditions where relevant.

A high-quality certified gas cannot compensate for an uncontrolled dilution system.

NABL 138 links dilution-system traceability to NABL 142, Policy on Metrological Traceability of Measurement Results. NABL 142 explains why metrological traceability is central to confidence in testing and calibration results.

Why gas dilution is a metrology issue

Suppose a laboratory has a certified gas with a relatively high concentration but needs five lower calibration points. A gas diluter combines the reference gas with zero gas to create those points.

If the flow ratio is wrong, every generated concentration can be biased. The analyzer may seem to respond correctly even though it is being challenged with a concentration different from the value assigned by the laboratory.

The dilution process must therefore be traceable, technically controlled, and supported by an appropriate uncertainty evaluation. Assessing only the gas cylinder is not enough.

Reference approaches for common gas analyzers

NABL 138 identifies reference approaches for several common analyzer types.

AnalyzerTypical reference approach
NOx analyzerCertified NO reference gas
CO analyzerCertified CO reference gas
SO₂ analyzerCertified SO₂ reference gas
O₃ analyzerReference ozone photometer
Ammonia analyzerCertified gas appropriate to the method
Benzene analyzerCertified benzene gas

Laboratories should use the current NABL 138 document and any applicable calibration or measurement standards when defining the exact method.

Why ozone calibration is different

Ozone is not usually treated in the same way as a stable bottled calibration gas. It must generally be generated and measured through a controlled reference system.

NABL 138 identifies a reference ozone photometer for O₃ analyzer calibration. The laboratory must therefore demonstrate competence in operating the ozone-generation and reference-measurement chain, not merely in reading the analyzer under calibration.

Potential influences include ozone-generator stability, gas flow, transfer-line materials, photometer performance, environmental conditions, and stabilization time.

How are PM2.5 and PM10 analyzers calibrated?

Particulate analyzers present a different problem from gas analyzers. They measure particle concentration using systems that may depend on sampled air volume, attenuation response, environmental sensors, and instrument-specific calculations.

NABL 138 includes PM2.5 and PM10 analyzers based on beta attenuation. Relevant reference equipment may include:

  • traceable attenuation discs or a suitable beta-attenuation calibration system;
  • a reference flow meter;
  • temperature sensors where applicable;
  • pressure sensors where applicable.

The criteria specify a minimum of three traceable discs for attenuation-coefficient calibration in this context.

PM monitor calibration should address the complete measurement chain rather than only the detector.

Why flow calibration matters for PM monitoring

Particulate concentration is related to the amount of particle mass collected or detected and the volume of air sampled.

If the monitor’s flow rate is wrong, the reported mass concentration can also be wrong—even when the particle-detection system is operating as intended.

A suitable PM calibration program may need to assess:

  • sample flow;
  • pressure;
  • temperature;
  • attenuation response;
  • reference-disc values;
  • sampling components;
  • instrument calculations and corrections;
  • software configuration;
  • repeatability.

This is why a detector check alone may not establish the performance of the full particulate-monitoring system.

NABL accreditation and CPCB requirements are different

NABL accreditation and Central Pollution Control Board requirements serve different purposes.

NABL accreditation demonstrates that a calibration laboratory is competent to perform the activities listed in its approved scope. CPCB requirements address environmental monitoring, station infrastructure, instrument specifications, and regulatory needs.

A NABL-accredited calibration certificate does not automatically prove that an analyzer complies with every CPCB specification. In the same way, an analyzer that meets a procurement or equipment specification still needs proper calibration and quality control.

The claim “NABL calibration means CPCB approval” should be avoided. These are separate forms of evidence.

What does metrological traceability mean?

Metrological traceability connects a measurement result to a recognized reference through a documented, unbroken chain of calibrations, with each step contributing to measurement uncertainty.

In air-quality equipment calibration, that chain may involve:

  • certified gas standards;
  • national metrology institutes;
  • accredited reference-material producers;
  • reference ozone photometers;
  • flow standards;
  • temperature and pressure standards;
  • attenuation references;
  • calibrated dilution equipment.

Traceability is not simply the possession of a certificate. The laboratory must determine whether the reference is suitable for the intended work.

That assessment should consider whether:

  • the reference remains within its valid calibration or certification period;
  • its range covers the intended calibration points;
  • its uncertainty is suitable;
  • the reference material is stable;
  • storage conditions are controlled;
  • the dilution or transfer process has been validated;
  • the complete traceability chain can be reconstructed.

Measurement uncertainty in air-quality calibration

Every calibration result has uncertainty. The laboratory must identify the significant contributors, quantify them where possible, and combine them through a technically justified uncertainty budget.

Calibration areaPossible uncertainty contributors
Reference gasCertified value, stability, storage, and reference-material uncertainty
Gas dilutionFlow ratio, controller performance, zero-gas purity, leakage, and repeatability
Analyzer responseResolution, repeatability, zero drift, span behavior, and stabilization
PM flowReference flow-meter uncertainty and temperature or pressure corrections
Attenuation systemReference-disc value, positioning, and analyzer response
On-site workEnvironmental conditions, equipment transport, setup, and field stability

A laboratory should build its uncertainty budget around its own equipment and method. Copying an uncertainty value from another laboratory is not enough, even when both laboratories calibrate the same analyzer model.

Calibration versus routine zero and span checks

Continuous monitoring stations often run automatic or manual zero and span checks. These checks are useful, but they are not automatically equivalent to a complete calibration.

Routine zero/span checkTraceable calibration
Tracks changes in analyzer responseEstablishes response against suitable references
May use an operational check standardUses controlled, traceable reference standards
Supports routine station quality controlSupports documented measurement performance
May cover only zero and one span pointMay assess multiple points across the required range
Helps identify driftQuantifies error and associated uncertainty

Routine checks can reveal that an instrument has changed. Calibration determines how its response compares with suitable reference values.

Preparing for NABL 138 accreditation

A laboratory seeking NABL 138 accreditation should begin with a realistic definition of its proposed scope. It should identify the analyzer types, pollutants, ranges, methods, and facility modes it can support with appropriate personnel and equipment.

The preparation process should include:

  1. Review the current NABL 138. Use the issue and amendment shown in the current NABL register.

  2. Define a technically supportable scope. Avoid applying for analyzers or ranges that the laboratory cannot calibrate with suitable references.

  3. Establish metrological traceability. This may require certified gases, reference photometers, flow standards, environmental sensors, and particulate references.

  4. Validate the calibration methods. Include gas generation, dilution, stabilization, data collection, calculations, and field procedures where applicable.

  5. Evaluate measurement uncertainty. Build budgets around the actual equipment and method.

  6. Demonstrate personnel competence. Staff should understand the measurement principle, reference system, calibration procedure, gas safety, data analysis, and uncertainty evaluation.

  7. Control relevant environmental conditions. Establish limits and records based on their potential effect on results.

  8. Create on-site and mobile procedures. Address equipment transport, field setup, environmental influences, and record control.

  9. Monitor the validity of results. Use suitable quality-control checks, comparisons, and other technical evidence.

  10. Use the current application documents. The NABL document register should be checked before submission because forms and amendments can change.

Personnel competence must be instrument-specific

General calibration experience does not automatically establish competence in air-quality analyzer calibration.

A technician who can calibrate a temperature indicator may not be qualified to calibrate a chemiluminescence NOx analyzer, an ozone photometer, or a beta-attenuation PM monitor.

Relevant competence may include:

  • knowledge of the analyzer’s measurement principle;
  • safe handling of calibration gases;
  • operation of dilution and gas-delivery systems;
  • flow measurement;
  • leak testing;
  • analyzer diagnostics;
  • data logging and calculations;
  • uncertainty evaluation;
  • management of field conditions;
  • recognition of unstable or invalid results.

Competence should be demonstrated through records and practical evidence, not job titles alone.

Common air-quality calibration mistakes

Several mistakes can undermine an otherwise well-documented calibration:

  • using certified gas with an unverified dilution system;
  • treating a zero/span check as a full calibration;
  • ignoring sample-flow calibration for a PM analyzer;
  • using expired or unsuitable reference gases;
  • overlooking transport effects during on-site calibration;
  • assigning the same uncertainty to every analyzer and range;
  • assuming CPCB compliance proves NABL calibration competence;
  • applying for an overly broad scope without the necessary references;
  • failing to record environmental conditions that affect the result.

Most of these errors share the same cause: focusing on one instrument while ignoring the rest of the measurement chain.

What customers should verify before ordering calibration

Before placing an order, customers should confirm that:

  • the laboratory’s NABL accreditation is current;
  • the analyzer type appears in its approved scope;
  • the required pollutant and measurement range are covered;
  • the approved facility mode matches the work required;
  • the laboratory’s calibration and measurement capability is suitable;
  • the proposed method meets the monitoring program’s needs;
  • the certificate will clearly report the results and relevant references.

A NABL logo or general claim of accreditation is not enough. The current scope should match the service being offered.

Why calibration matters in Delhi-NCR and other heavily monitored regions

Regions with dense ambient-air monitoring networks rely on measurements from many analyzers operating across different locations and conditions.

The usefulness of that network depends on comparability. If two instruments respond differently to the same pollutant concentration, apparent differences between locations may reflect measurement bias rather than actual air quality.

Calibration supports consistency across the network. It helps monitoring organizations compare results, identify trends, assess equipment performance, and maintain confidence in reported concentrations.

The central principle: air-quality data need a reliable measurement chain

Behind every reported pollutant concentration sits a chain that may include:

  • sample collection;
  • gas delivery or airflow;
  • analyzer response;
  • reference standards;
  • temperature and pressure corrections;
  • software calculations;
  • routine quality-control checks;
  • traceable calibration.

NABL 138 addresses the competence of laboratories responsible for a critical part of that chain.

Reliable air-quality decisions start with reliable measurements. Reliable measurements start with suitable references, controlled methods, competent personnel, defensible uncertainty, and traceability that can be demonstrated.

Frequently Asked Questions

What is NABL 138?

NABL 138 is the Specific Criteria for Air Quality Monitoring Equipment Calibration Laboratories. The NABL register lists Issue 01 with Amendment 02 dated 3 November 2025.

Which instruments are covered by NABL 138?

The criteria cover gas analyzers for pollutants including NOx, CO, SO₂, ozone, ammonia, and benzene, along with PM2.5 and PM10 analyzers.

How are gas analyzers calibrated?

Gas analyzers are exposed to known reference concentrations. The laboratory compares the instrument’s response with those values using certified gases or appropriate reference systems, controlled gas delivery, traceable dilution where required, and a defined uncertainty evaluation.

How are PM2.5 and PM10 analyzers calibrated?

For beta-attenuation analyzers, calibration may involve traceable attenuation references or a suitable calibration system, a reference flow meter, and temperature or pressure measurements where applicable.

Is a zero/span check the same as calibration?

No. A zero/span check helps monitor routine analyzer performance. A full calibration establishes instrument response against suitable traceable references and reports the associated results and uncertainty.

Does NABL calibration mean an analyzer is CPCB approved?

No. NABL accreditation demonstrates laboratory competence for an approved calibration scope. CPCB specifications and regulatory requirements serve a different purpose.

Is ISO/IEC 17025:2017 still current in 2026?

Yes. ISO states that ISO/IEC 17025:2017 was reviewed and confirmed in 2023.

Official references

Editorial note: Accreditation documents, application forms, technical specifications, and regulatory requirements can change. Laboratories should consult the current NABL, ISO, CPCB, and applicable regulatory documents before defining a calibration scope or offering a service.