NABL & ISO/IEC 17025:2017 Calibration Reference Portal
Official Lab Services: multitek.in
Solutions for: Quality Heads & Auditors Shop Floor Inspectors Tool Room Technicians Engineering Students

How to Verify NABL Torque Calibration Scope


How to Verify a NABL Torque Calibration Scope Before Ordering

"NABL accredited" is useful information, but it is not the end of the purchasing check. Accreditation applies to a defined scope, and the scope needs to match the calibration you are ordering.

Start with the correct accreditation framework

NABL FAQ states that NABL provides accreditation to calibration laboratories in accordance with ISO/IEC 17025. ISO/IEC 17025:2017 is the international standard used for assessing laboratory competence, impartiality and consistent operation.

The practical purchasing question is therefore not simply, "Is the laboratory accredited?" It is, "Is the calibration I need included within the laboratory's current accredited scope?"

Check the discipline and instrument type

Locate the laboratory's current scope and find the relevant mechanical or torque activity. Confirm that the wording covers the type of instrument you are sending, such as a torque wrench or other torque-generating device.

Do not assume a laboratory accredited for force, pressure or another mechanical quantity is automatically accredited for torque.

Check the measurement range

The accredited range should cover the torque values required for your tool and intended use. A laboratory may have excellent capability at one torque range but not cover your full wrench range under accreditation.

For example, if your controlled work uses a wrench around 300 N·m, an accredited scope that ends far below that value does not cover the actual need.

Check the CMC information

NABL scopes can include a Calibration and Measurement Capability, often abbreviated CMC. This is important when your acceptance decision requires sufficiently small measurement uncertainty.

The correct question is not "Which laboratory has the smallest number?" in isolation. The capability should be suitable for your required range and decision risk.

Check the method and any scope conditions

Read the scope entry carefully for the stated method or procedure and any limitations. If your purchase order specifies a particular method or standard, confirm that the requested service and the accredited scope are aligned.

ISO 6789-2:2017 is the published ISO standard covering calibration of hand torque tools and determination of measurement uncertainty.

Verify current status, not an old PDF

Accreditation status and scope can change. Use the current NABL information rather than relying on a certificate or scope PDF saved years ago.

NABL Accreditation Process and Scope explains that accreditation is granted within defined scopes. Recheck status when placing important or recurring orders.

Put the requirement on the purchase order

If accredited calibration is mandatory, state it explicitly. A useful purchase order can specify:

  • instrument and range;
  • required calibration method where applicable;
  • NABL-accredited calibration within scope;
  • as-found results before adjustment where required;
  • measurement uncertainty reporting;
  • serial number and certificate identification;
  • approval before repair or adjustment if needed.

This reduces the chance that the tool returns with a valid certificate that does not meet the actual customer or quality-system requirement.

Frequently Asked Questions

Does NABL accreditation cover every calibration a laboratory offers?

No. Accreditation applies to the activities included in the laboratory's defined scope.

Why should I check the range on the NABL scope?

Because the accredited capability must cover the relevant torque values for the instrument and intended use.

What is CMC on a calibration scope?

CMC means Calibration and Measurement Capability. It describes the laboratory's accredited measurement capability under stated conditions and is relevant when evaluating whether the uncertainty is suitable for your need.

Verify the scope before sending the wrench

Match instrument type, range, method, CMC and current accreditation status to the purchase requirement.

Authoritative Sources

Technical note: Always use current NABL scope information. Accreditation status and scope can change over time.