What does the FAA
actually require?
For a U.S. certificated Part 145 repair station, 14 CFR §145.109(b)
requires test and inspection equipment and tools used to make airworthiness
determinations to be calibrated to a standard acceptable to the FAA.
Section 145.211 separately requires the repair station's quality-control manual
to describe how measuring and test equipment is calibrated, including the
intervals used. The regulations do not establish one universal annual calibration
interval. [1]
FAA AC 145-9A expands on the practical control system. It recommends
documented traceability, a controlled equipment list, defined recall
responsibility, calibration records and identification that prevents
inadvertent use of uncalibrated equipment. It also recognizes several potential
traceability routes rather than stating that every calibration must be
performed directly by NIST. [21]
What does EASA
Part-145 require?
EASA 145.A.40 requires tools and equipment, particularly test equipment
as appropriate, to be controlled and calibrated according to an officially
recognized standard at a frequency sufficient to ensure serviceability and
accuracy. Calibration records and traceability to the standard used must be
retained. The associated AMC says manufacturer instructions normally provide the
interval unless results support a different period. [22]
What does “traceable” really mean?
Metrological traceability is not simply a
“NIST traceable” statement on a certificate. NIST describes it as a measurement
result linked to a specified reference through a documented, unbroken chain of
calibrations, with each link contributing to measurement uncertainty. NIST also
warns that traceability by itself does not prove a result is fit for a
particular measurement purpose. [23]
Copyable compliance
checklist
|
Check |
Organization entry |
|
Regulatory jurisdiction and approval basis |
|
|
Is the tool used to make an airworthiness determination? |
Yes / No |
|
Unique tool ID / serial number recorded |
|
|
Calibration standard accepted by organization/authority |
|
|
Traceability evidence reviewed |
|
|
Calibration interval and basis documented |
|
|
Current calibration date |
|
|
Next due date |
|
|
Calibration record/certificate reference |
|
|
Point-of-use status identified |
|
|
Overdue-tool control defined |
|
|
Out-of-tolerance response defined |
|